EasyManua.ls Logo

SICK DUSTHUNTER T100

SICK DUSTHUNTER T100
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Calibration
17
5 Calibration
For a stable and accurate data acquisition, periodical calibration is necessary.
There are two calibration methods; automatic calibration (AIC) and manual calibration.
Automatic calibration (AIC)
By the set time span or from external contact input, the zero and span calibration is
automatically performed. Pressing [AIC] on the touch panel also enables to perform
automatic calibration.
Reference
Refer to " 5.3 AIC " (page 22).
Manual calibration
A calibration method given by manual operation with flowing the calibration gas
voluntarily.
Reference
Refer to " 5.4 Manual Calibration " (page 28).
Note
In calibration, the concentration output is held according to the specification.
5.1 Calibration Screen (CAL. screen)
The CAL. screen is the basic screen to perform calibration.
Pressing [CAL.] in the MEAS. screen 1 (Measurement value display) (page 14 Fig. 7) displays
the CAL. screen.
Fig. 11 CAL. screen
Calibration mode
Range
Function key
Alarm/Caution

Table of Contents

Related product manuals