ASSEMBLY AND INSTALLATION 3
Onducts with strongly reflective material we recommend to define the measurement axis
according to the following figure to prevent disturbing reflections influencing the measured
value.
Fig. 17: Measurement axis in in strongly reflective ducts
Work to be performed
▸ Measure the fitting location and mark the assembly location.
Leave enough clearance around the flange with tube to fit the sender/receiver unit and
reflector.
Fig. 18: Clearance for sender/receiver unit and reflector (dimensions in mm)
▸ Remove insulation (when fitted)
< 90 °
< 90 °
Strongly reflective duct wall
Measurement axis
TOP
400 400
600 300 200
Flange with tube
Duct wall