7ML19981FB05 MultiRanger 100/200 – INSTRUCTION MANUAL Page 7
mmmmm
Specifications
Cable
Do not use coaxial cable for transducer (see General Appendix F: Upgrading on
page 246 for more information)
transducer and mA output signal to be 2 copper conductors, twisted with shield/drain
wire, 300 Vrms, 0.5 mm
2
(22-18AWG), nominal capacitance between adjacent
conductors @ 1kHz = 62.3 pF/m (19 pF/ft). Nominal capacitance between conductor
and shield @ 1kHz = 108.3 pF/m (33 pF/ft) (Belden®
1
8760 is acceptable)
365 m maximum
1.
Belden is a registered trademark of Belden Wire & Cable Company.
Note: The MultiRanger is to be used only in the manner outlined in this instruction
manual or protection provided by the equipment may be impaired.