EasyManua.ls Logo

Siemens SIPROTEC 5 - Settings

Siemens SIPROTEC 5
490 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter: Temperature unit
To change the display and evaluation of measured temperature values from °C to °F, adapt the DIGSI user
default settings accordingly.
Proceed as follows:
In DIGSI select the menu item Extras --> Settings.
In the Settings view select the menu item DIGSI 5 User preferences.
Under Standard unit system change the setting value of the unit system used from
SI units
to
US
units
.
[scfahrht-190214-01, 1, en_US]
Figure 6-59 Change of the Display Between °C and °F
The following settings and information table shows only 1 of the 12 sensors, as the setting possibilities of the
12 sensors do not differ.
Settings
Addr.
Parameter C Setting Options Default Setting
General
_:2311:103 General:Port
port F
port E
port P
port N
port J
port J
_:2311:106 General:Slave address 1 to 254 1
Sensor 1
_:11611:102 Sensor 1:Sensor type
Pt 100
Ni 100
Ni 120
Pt 100
Sensor 2
_:11612:102 Sensor 2:Sensor type
Pt 100
Ni 100
Ni 120
Pt 100
6.5.7.7
Function-Group Types
6.5 Function-Group Type Analog Units
200 SIPROTEC 5, Fault Recorder, Manual
C53000-G5040-C018-5, Edition 11.2017

Table of Contents

Other manuals for Siemens SIPROTEC 5

Related product manuals