EasyManua.ls Logo

Smartrise C4

Default Icon
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C4 User Manual
Page 254 ©2021 Smartrise Engineering, Inc. All Rights Reserved October 25, 2021
4. From the DEFAULT FLOORS menu, select if restoring original learned floors.
Figure 462: DEFAULT FLOORS? Menu
5. From the DEFAULT FLOORS? menu, select Yes to default floors.
NOTE: If not defaulting floors, select NO to back out.
Only the Learned floor values are defaulted as shown in the figure below.
Figure 463: DEFAULTING PARAMS Menu
6. When the system has completed defaulting all parameters, the DEFAULTING PARAMS
menu displays Parameters Defaulted as shown in the figure below.
Figure 464: DEFAULTING PARAMS Menu Parameters Defaulted
29.17.2 Default S-Curve
The user has the option restore the original S-Curve.
The following procedure describes how to select default S-Curve.
1. Navigate to MAIN MENU | SETUP | MISCELLANEOUS. See Figure 58.
2. From the MISCELLANEOUS menu, scroll and select Default. See Figure 460.
3. From the DEFAULT menu, scroll and select Default S-Curve.
Figure 465: DEFAULT Menu Default S-Curve

Table of Contents

Other manuals for Smartrise C4

Related product manuals