14 Probes and Biopsy
Basic User Manual 137
14.2.2 Assembling the Biopsy Bracket
■ To Assemble the Surface Biopsy Bracket
Biopsy Guide Tube
Fixing Bracket
Locking Screw
Tube Holder
Handle
Perform the following steps.
1. Check and ensure that there is no damage, deformation, malfunction, loose or missing parts on the biopsy
bracket before assembly.
2. Wear sterile gloves.
3. Unfold the probe sheath and apply an appropriate amount of coupling gel to the inside of the sheath and onto
the face of the probe.
4. Hold the probe and unroll the sheath onto the probe. Pull the probe sheath tightly over the face of the probe to
remove any wrinkles.
5. Align the biopsy bracket with the orientation mark of the probe.
Orientation
mark
6. Attach the biopsy bracket to the probe and fasten the biopsy bracket with the locking screw.