147
Appendix B EMC Guidance and
Manufacturer’s Declaration
B.1 Electromagnetic Emissions
The equipment is intended for use in the electromagnetic environment specied below. The customer or the user of
the EQUIPMENT should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment and Guidance
RF emissions
CISPR 11
Group 1
The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
icker emissions
IEC 61000-3-3
Complies
B.2 Electromagnetic Immunity
The equipment is intended for use in the electromagnetic environment specied below. The customer or the user of
the equipment should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment and
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV Air
±6 kV Contact
±8 kV Air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines;
±1 kV for input/output
lines
±2 kV for power
supply lines;
±1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
±2 kV line to earth
Mains power quality should be that
of a typical commercial or hospital
environment.