3 Preparing the System
20 Basic User Manual
3.2 System Moving/Positioning
Leave at least 20cm at the back and both sides of the system for ventilation. Otherwise, the temperature rise could
cause failures.
Unlock
Lock
Brake
Perform the following steps to position the system.
1. Unlock the four foot brakes.
2. Move the system with the handle.
3. Lock the four foot brakes after placing it in the desired location.
3.3 Connecting and Powering the System
Two methods can be used to supply power for the system: mains supply and built-in rechargeable lithium battery.
When the adapter is connected to the mains supply, the battery is charged until it is full.
3.3.1 Using the Mains Supply
Two switches are provided for the system.
The power switch located on the peripheral device panel is used to connect or disconnect the system with the
mains supply.
The key on the side panel is used to start up or shut down the system.
!
Ensure the system works with the required voltage before you connect it to the mains supply. For details, refer
to Appendix A Specications.
Do not pull the power cable forcibly.
To connect the system:
1. Connect the power grounding cable to the grounding terminal.
2. Connect the power cable to the mains power socket on the peripheral device panel of the system.
3. Set the power switch on the peripheral device panel to the position.