EasyManua.ls Logo

Steris Reliance EPS Endoscope Processing System - Page 106

Steris Reliance EPS Endoscope Processing System
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-5
Cycle and Control Value Programming Operator Manual 920011-649
3. Press ENTER touch pad to confirm selection. Display moves on
to WASH segment of selected cycle.
Display shows, for example:
__ blinking
and printer records, for example:
==========================
HH:MM:SS XM MM-DD-YY
==========================
MODIFY: SINGLE
==========================
4. Press SCROLL arrows (left or right) to change time value of
WASH 1 segment (adjustable from five to ten minutes).
NOTE: If a value is out of range, display shows OUT OF
RANGE–SEE OPERATOR MANUAL for four seconds, then
control returns to current display.
5. Press ENTER touch pad to confirm new WASH 1 segment time
value. Printer records:
============================
CYCLE MODIFICATION SAVED
============================
NOTE: For new values to be saved, it is necessary to scroll through
all phases of a cycle and confirm values by pressing ENTER touch
pad after each phase, even if no value was changed.
- or -
To cancel a modification, press ESC touch pad once. Display
returns to cycle being modified. Values are not saved. Printer
records:
============================
CYCLE MODIFICATION NOT
SAVED
============================
6. Display moves on to AIR PURGE phase time of selected cycle.
Display shows:
__ blinking
WASH
RECIRC MM:SS
AIR PURGE
RECIRC 0
4:00

Table of Contents

Related product manuals