5-6
920011-649 Operator Manual Cycle and Control Value Programming
and printer records, for example:
==========================
HH:MM:SS XM MM-DD-YY
==========================
MODIFY : SINGLE
==========================
7. Press SCROLL arrows (left or right) to change time value of AIR
PURGE phase (adjustable from 4 to 30 minutes).
IMPORTANT:
A decrease in air purge efficiency may be caused
by a clogged HEPA filter or pre-filter. Refer to
S
ECTION 7,
T
ROUBLESHOOTING
, before changing air purge phase time.
NOTE: If a value is out of range, display shows OUT OF
RANGE–SEE OPERATOR MANUAL for four seconds, then
control returns to current display.
8. Press ENTER touch pad to confirm new AIR PURGE phase time
value. Printer records:
============================
CYCLE MODIFICATION SAVED
============================
NOTE: For new values to be saved, it is necessary to scroll
through all phases of a cycle and confirm values by pressing
ENTER touch pad after each phase, even if no value was
changed.
- or -
To cancel a modification, press ESC touch pad once. Display
returns to cycle being modified. Values are not saved. Printer
message is:
==========================
CYCLE MODIFICATION NOT
SAVED
==========================
9. When modifications are completed for selected cycle, supervisor
may continue to modify values for next cycle, or exit Change
Values menu by pressing ESC touch pad. Display shows:
__ blinking
10. Press SCROLL arrows (up or down) to choose between YES or
NO.
11. Press ENTER touch pad to confirm selection.
PRINT CYCLE
YES