5-7
Cycle and Control Value Programming Operator Manual 920011-649
If YES is selected, printer records values of cycle just modified,
for example:
===========================
HH:MM:SS XM MM-DD-YY
===========================
CYCLE - SINGLE
WITH WASHING
===========================
WASH
WATER TYPE HOT
TEMPERATURE 48.0 C
RECIRC. TIME 05:00
CHEMICAL KLENZYME
CONCENTR. 2 mL/L
RINSE
WATER TYPE HOT
RECIRC. TIME 00:40
HLD
WA
TER TYPE HOT
TE
MPERATURE 50.0 C
GENERATION 04:00
DISINFECTION 06:00
RINSE 1
WATER TYPE HOT
RECIRC. TIME 00:40
RINSE 2
WATER TYPE HOT
RECIRC. TIME 00:40
AIR PURGE
RECIRC. TIME 04:00
===========================
Display shows:
__ blinking
12. Press SCROLL arrows (up or down) to choose another
Supervisor mode menu.
- or -
Press ESC touch pad to exit Supervisor mode.
SCROLL TO MODIFY
SINGLE