7-7
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 07
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 07 =
SUMP DRAIN TOO LONG
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 08
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 08 =
CIP 200 EMPTY
1. CIP 200 Acid-Based Process and Research Cleaner Bag-
in-a-Box container is empty.
• Replace Bag-in-a-Box container (refer to S
ECTION 6.8).
2. If container is not empty, cleaner injection system is
clogged. If container was just changed, it may be
improperly connected.
• Wearing appropriate Personal Protective Equipment
(PPE), clean Bag-in-a-Box injection system, and reinstall
Bag-in-a-Box container (refer to S
ECTION 6.8).
3. Defective cleaner pressure switch, call STERIS.*
Alarm sounds, and display shows:
FAULT 09
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAUL
T 09 =
KLENZYME EMPTY
1. Klenzyme Enzymatic Presoak and Cleaner Bag-in-a-Box
container is empty.
• Replace Bag-in-a-Box container (refer to S
ECTION 6.8).
2. If container is not empty, cleaner injection system is
clogged. If container was just changed, it may be
improperly connected.
• Wearing appropriate Personal Protective Equipment
(PPE), clean Bag-in-a-Box injection system, and reinstall
Bag-in-a-Box container (refer to S
ECTION 6.8).
3. Defective cleaner pressure switch, call STERIS.*
FAULT 10 – Not used.
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION