7-8
920011-649 Operator Manual Troubleshooting
Alarm sounds, and display shows:
FAULT 11
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 11 =
DOOR OPEN
1. If door opens during cycle, call STERIS.*
2. If door is closed, resume cycle by pressing START touch
pad. If situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 12
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 12 =
SUMP RTD HEATING MALFUNCTION
1. Verify hot tap water (HTW) supply temperature as per
equipment drawing (920-011-656).
2. Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 13
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 13 =
SUMP RTD DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
FAULT 14 – Not used.
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION