EasyManua.ls Logo

Steris Reliance EPS Endoscope Processing System - Page 146

Steris Reliance EPS Endoscope Processing System
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-9
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 15
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 15 =
AIR PURGE RTD DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
FAULT 16 – Not used.
FAULT 17 – Not used.
FAULT 18 – Not used.
Alarm sounds, and display shows:
FAULT 19
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 19 =
BOARD RTD OVERHEAT
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION

Table of Contents

Related product manuals