Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 43
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 43 =
SINGLE/DOUBLE CYCLE OVER TEMPERATURE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
FAULT 44 – 50 Hz Processors. Refer to Operator Manual - Information Supplement For
50 Hz Processors (P920501-322) for specific information.
FAULT 45 –
50 Hz Processors. Refer to Operator Manual - Information Supplement For
50 Hz Processors (P920501-322) for specific information.
Alarm sounds, and display shows:
FAULT 46
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAUL
T 46 =
LEAK TEST PORT 1: OVERPRESSURE
Press ALARM REPLY touch pad to silence audible alarm.
Pressure inside endoscope 1 over maximum 4.33 psig
(0.30 bar).
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Possible cause:
Leak test port 1 pressure transducer not calibrated.
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION