7-18
920011-649 Operator Manual Troubleshooting
Alarm sounds, and display shows:
FAULT 40
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 40 =
LEAK TEST PORT 1: PRESSURE SENSOR DEFECTIVE
Press ALARM REPLY touch pad to silence audible alarm.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Alarm sounds, and display shows:
FAULT 41
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 41 =
LEAK TEST PORT 2: PRESSURE SENSOR DEFECTIVE
Press ALARM REPLY touch pad to silence audible alarm.
Cycle aborted; users are required to restart a processing
cycle. If unsuccessful, call STERIS.*
Alarm sounds, and display shows:
FAULT 42
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 42 =
FILLING WATER TEMPERATURE TOO HIGH
1. Verify hot tap water (HTW) supply temperature is as per
equipment drawing (920-011-656). Filling water
temperature, as read by sump RTD, can be verified by
pressing SINGLE
or DOUBLE touch pad twice during
filling phase of a cycle.
2. Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.