7-17
Troubleshooting Operator Manual 920011-649
Alarm sounds, and display shows:
FAULT 37
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 37 =
PROTECTION RTD DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 38
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 38 =
FRESH DRY CHEMISTRY CONTAINER NOT DETECTED
A new Reliance dry chemistry container was not inserted in
Reliance CDS prior to cycle start. Cycle is aborted.
Start new cycle after ensuring a new Reliance dry
chemistry container is installed in Reliance CDS and
ensuring latch is closed on Reliance CDS. If situation
reoccurs, call STERIS.*
Alarm sounds, and display shows:
FAULT 39
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 39 =
DRY CHEMISTRY CONTAINER DETECTION
DEFECTIVE
Try to resume cycle by pressing START touch pad. If
situation reoccurs, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
FAULT POSSIBLE CAUSE AND CORRECTION