3. Remove the four Allen screws from the
Service Head using a 5mm Allen wrench.
4. Locate the cable kit end and pass through
the service head opening.
0682-001-345 Wedge Kit
MMP200 only
Note e Shelf Wedge kit has a
thick and a thin component.
Orientation of the thin side
towards the front of the ser-
vice head is important.
5. Verify which Wedge is the thinner wedge
of the 2 halves of the Wedge Kit
6. Place the thinner Wedge towards the front
of the Service Head
7. Place the thicker Wedge towards the rear
of the Service Head.
8. Connect Service Head to articulating arm.
9. Tighten the four Allen screws to 14.75 lb-
(20 Nm) using a 5mm Allen wrench.
10. s