6. Adjust friction brakes (Section 11.2.1.4) until the suspension no longer dris.
11.3 Adjusting the Cardanic Suspension
Use a 4mm Hex key to adjust the brake screws on the Cardanic suspension. If the brakes are
adjusted properly, the light head will not tilt when released and should be easily positioned with the
light handle.
• To increase the brake force, turn the screw clockwise.
• To decrease the brake force, turn the screw counterclockwise.
Brake Screws
Cardanic Suspension Brake Screws
(e Low Ceiling light only has one brake screw)