14. Install the in-light camera.
15. Adjust tension in the Spring Arm to compensate for additional weight of the newly installed
camera.
Aer performing test and adjustments, reinstall the cover on the light.
Warning Failure to tighten the three thumb screws may result in the In-Light Cam-
era falling out during use!
13.1.6 Testing
1. Turn the light on and verify that it functions normally and the wall touch panel controls the
light.
2. Connect a monitor to the video output of the power supply box.
3. Allow the light to warm up for at least 15 minutes.
4. While monitoring the image quality, perform the following adjustments:
a. Rotate Luminance Carrier level adjust-
ment potentiometer P1 clockwise until
the image is lost.
b. Slowly rotate P1 counterclockwise until
the image returns and is stable. Rotate
1/8th of a turn further counterclockwise.
c. Rotate Chrominance Carrier level adjust-
ment potentiometer P2 clockwise until
the color signal is lost and the image is
black and white.
d. Slowly rotate P2 counterclockwise until
the color returns to the picture. Rotate
1/8th of a turn further counterclockwise.