EasyManua.ls Logo

Zonge GDP-32 - Sample Rates and Antialias Filter Delays

Default Icon
419 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDP-32
II
INSTRUCTION MANUAL
October 2002 Section 12, Page 18
12.7 SAMPLE RATES AND ANTIALIAS FILTER DELAYS
FREQ SAMPLE SAMP'S NYQUIST LOW PASS FILTER SAMPLE
Fo RATE /CYCLE Fn Fc Delay INTERVAL
32 32768 1024 16384 24.8 KHz 26 µs 30.5 µS
16 32768 2048 16384 24.8 26 µs 30.5 µS
8 32768 4096 16384 24.8 26 µs 30.5 µS
4 32768 8192 16384 24.8 26 µs 30.5 µS
2 16384 8192 8192 12.4 39 µs 61.0 µS
1 8192 8192 4096 6.4 64 µs 122.1 µS
.5 4096 8192 2048 3.2 118 µs 244.1 µS
.25 2048 8192 1024 1.6 226 µs 488.3 µS
.125 1024 8192 512 0.8 440 µs 976.6 µS
.0625 512 8192 256 0.4 880 µs 1,953 µS