GDP-32
II
INSTRUCTION MANUAL
May 2002 Section 3, Page 2
3.1 INITIAL RECEIVER CHECK
Upon first receiving the GDP-32
II
system, perform a brief check to verify that the receiver is
ready for operation. Remove the receiver from its shipping case and inspect it for exterior
damage due to shipment. If there is anything more than cosmetic damage contact Zonge
Engineering immediately.
LARGE CASE GDP-32
II
(UP TO 16 CHANNELS)
Case Bottom Access
1. Lay the receiver on its back and open the clasps to the bottom battery compartment.
2. Carefully open the bottom compartment.
3. Remove the 12 screws from the bottom panel of the main case. Make sure all cables
are attached, and press in each card to make sure they are fully seated in their bus
connectors.
NOTE: All 16-channel GDP-32
II
’s are pre-wired for 16 channels, regardless of how
many channels were ordered. After checking attachments, replace the bottom panel,
making sure that no wires are being pinched. Secure the panel with 12 screws. Tighten
firmly, but do not over tighten.
4. With the battery compartment still open, connect the battery pack.
5. Check the main fuse holder and make sure that there are spare fuses in the battery
compartment.
6. Close the battery compartment and secure with the retaining clasps.
Front Panel Access
1. Set the receiver upright and open the top lid.
2. Unscrew the 12 screws from the top panel and carefully remove the panel (be careful
to not stress the connecting wires).
3. Verify that each board is firmly mounted and that all cables are firmly connected to
the appropriate boards; there should be no unconnected cables.
4. Replace the top panel being careful to not crimp any cables or wires and replace the
12 securing screws.
NOTE: Turn the screws until the head touches the top of the panel, then turn it ¼ turn
more.