4-57 2000-OSM, F1
6. Cursor to the Function field and scroll to Pres Contrl. In Figure 4-49, with Pres Contrl displayed,
the Value column lists Two as the function to be performed.
***** Method Table #01 *****
Time Function Value
[0001] Pres Contrl Two
0010 Pres Check One
0120 Noise Calc On
0125 Noise Calc Off
0128 Auto Zero N/A
0130 Skp, Nxt,Str= Nine
0132 Valve On One
0155 Valve Off One
More Below
Insert Delete Cal Define
Exit Line Line & Cycle
Time
F1 F2 F3 F4
Figure 4-49. METHOD TABLE #01 SCREEN
7. Move the cursor to the Value column and the Pressure Control Table appears, with the
cursor on Entry line 02 (see Figure 4-50). Note the Entry line number when you enter the table so
you can exit the table from the same line, if desired.
***** Pressure Control *****
Entry Zone Set Point Ramp Rate
01 Pres Zone1 150.0 25.00
[02] Pres Zone1 250.0 50.00
03 Pres Zone1 100.0 01.00
04 Pres Zone1 0.000 0.000
05 Pres Zone1 0.000 0.000
06 Pres Zone1 0.000 0.000
07 Pres Zone1 0.000 0.000
08 Pres Zone1 0.000 0.000
More Below
Exit
F1 F2 F3 F4
Figure 4-50. PRESSURE CONTROL SCREEN