1-6 2000-OSM, F1
Gas Control (Digital)
Control Method: Closed loop PID; temperature stabilized
Number of Zones: 1 to 5
Filtration: 2 microns at inlet, provided
Inlet Pressure:
Minimum: Setpoint + 10 psig (69 kPa, 0.7 bar)
Maximum: 100 psig (689 kPa, 6.89 bar)
NOTE
It is the customer's responsibility to set relief devices used to protect the analyzer
and prevent hazardous releases at a maximum of 200 psig (1380 kPa, 13.80 bar).
Range: 0 to 100 psig (0 to 689 kPa, 0 to 6.89 bar), bubble tight, non-
venting
Gauges: Electronic readout: 0.01 psig (0.07 kPa, 0.7 mbar) resolution
Setpoint resolution: 0.01 psig (0.07 kPa, 0.7 mbar)
Temp Coefficient: None
Regulation: See below
Accuracy:
0 to 50 psig: 1.7%
50 to 100 psig: 2.7%
Repeatability: ±0.1 psig (±0.7 kPa, ±7 mbar)
Allowable Gasses: H
2
, He, N
2
, Air, Ar
Quality: GC Grade
Flow Adjustment: Oven mounted needle valves or flow controllers with external
adjustment
Tube Fittings: 316 SS Gyrolok (standard)
316 SS Swagelok (optional)
1/16, 1/8, 1/4 inch connections