EasyManuals Logo

ABB PGC2000 User Manual

ABB PGC2000
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #259 background imageLoading...
Page #259 background image
2000-OSM, F1 8-5
Disassembly on the Analyzer
See Figure 8-4 for typical sample valve component location.
Figure 8-4. LIQUID SAMPLE VALVE COMPONENT LOCATION
1. Stop the analysis at the end of a cycle.
2. Turn off power, carrier, sample, and air to the analyzer.
3. Open the isothermal oven door.
4. Label all tubing connections to the Liquid Sample Valve.
5. Disconnect all tubing from the Liquid Sample Valve.
6. Mark the analyzer and flange to ensure the flange is reinstalled in the correct orientation.
7. Remove the four screws holding the flange to the side of the analyzer.
8. Separate the LSV from the analyzer, being careful not to stress the vaporizer assembly cabling.
9. Remove the insulation cover and insulation from around the vaporizer assembly.
10. Remove the four screws that retain the two parts of the vaporizer assembly.
11. Separate the vaporizer assembly from the vaporizer chamber.
12. Remove the two cone point set screws from the cylinder.
13. Slide the cylinder off the piston and body.
14. Mark on the body the location of the sample chamber.
15. Using a hex key wrench, rotate the sleeve counterclockwise until it is loose.
16. Unscrew the body from the flange and vaporizer chamber.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB PGC2000 and is the answer not in the manual?

ABB PGC2000 Specifications

General IconGeneral
BrandABB
ModelPGC2000
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals