7-26 2000-OSM, F1
5. Routine validation comparisons will help in the early detection of discrimination.
TP 33: CHECKING METHOD TABLE CONFIGURATION
Verify the method configuration tables as follows:
1. Press the F1 soft key until the Background screen is displayed (see Figure 7-11).
********** B A C K G R O U N D **********
ABB Analytical GCC Remote 0.1
803V029-1R (Std GCC) (C) 2001
**ALARM**
**ALARM** 18-JUN-2002 TUE 08:16:01
**ALARM**
Analysis Time :00114 Peaks Found:00
Cycle Time :00000 State:Idle
Current Stream :01 Key Component:
Purging Stream :01 none
Method :01 entered
Exit Manual Cur Report,
To Control Basic I/O & Alarms
Commands Options
F1 F2 F3 F4
Figure 7-11. BACKGROUND SCREEN
2. On the Background screen press the F1 (Exit to Commands) soft key.