2000-OSM, F1 3-5
Checking Streams
To verify the operation of each stream, perform a single analysis on each one, as follows:
1. On the Background screen press the F1 (Exit to Commands) soft key.
2. On the Commands screen cursor to ANALYSIS CONTROL, if necessary, and then press the F3
(Streams) soft key.
3. On the Stream Assignment screen, determine which streams are active. Then press the F1 (Exit)
soft key to return to the Commands screen.
4. On the Commands screen press the F2 (Request Analysis) soft key.
5. On the Request Analysis screen (see Figure 3-5), select the stream number to be analyzed.
6. Place the cursor on “Stream” and then press the F2 (Accept The Request) soft key.
7. Press the F1 (Escape) soft key to return to the Commands screen.
8. On the Commands screen, press the F4 (Start/Stop Analysis) soft key.
9. On the Start/Stop Analysis screen press the F2 (Start Analyzer) soft key. The controller will start
the analysis on this stream when all other requests have been completed.
10. Press the F1 soft key as necessary to return to the Background screen.
11. Watch the screen to determine when the selected stream has purged and is starting its analysis.
12. Press the F3 (Cur Report, Basic I/O, & Options) soft key.
13. On the Report, Basic I/O and Options screen, cursor to CURRENT REPORT, if necessary, and
press the F2 (View Current Report) soft key. The desired report will display at the end of the
analysis.
14. When the report is complete, press the F1 soft key as necessary to return to the Background
screen.
15. To repeat the process for remaining streams or to rerun this stream, repeat steps 1 through 14.
16. When checkout is complete press the F1 soft key to return to the Background screen.
ANALYTICAL OPERATION
Once you have verified the calibration data and each stream’s data, you are ready to place the
analyzer into full analytical operation.
Starting Analytical Operation
1. Display the Background screen.
2. Press the F1 (Exit to Commands) soft key.
3. On the Commands screen cursor to ANALYSIS CONTROL, if necessary, and press the F4
(Start/Stop Analysis) soft key.