2000-OSM, F1 7-27
3. On the Commands screen (see Figure 7-12) cursor to highlight TABLE EDIT. Then press the F2
(Methods) soft key.
****** COMMANDS ******
ANALYSIS CONTROL
REPORTS
TABLE EDIT
PRINTER
STORAGE & CONFIG.
NAMES
MISC.
TESTING/SETUP
SIM DISTILLATION
Go to Other
Background Methods Tables
F1 F2 F3 F4
Figure 7-12. COMMANDS SCREEN
4. When the Method Table #?? screen appears, enter “1” for the method number and then press the
F2 (Edit the Table) soft key.
5. The Cal Define & Cycle Time screen appears. Press the F1 (Continue) soft key to display the
first eight lines of Method Table # 01 (see Figure 7-13).
***** Method Table #01 *****
Time Function Value
0001 Temp. Contrl Two
0120 Noise Calc On
0125 Noise Calc Off
0128 Auto Zero N/A
0130 Skp, Nxt,Str= Nine
0132 Valve On One
0155 Valve Off One
0160 Distillation On
More Below
Exit Insert Delete Cal Define
Line Line & Cycle
Time
F1 F2 F3 F4
Figure 7-13. METHOD TABLE SCREEN