EasyManuals Logo

ABB PGC2000 User Manual

ABB PGC2000
400 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
7-28 2000-OSM, F1
6. Refer to the table listings provided with your chromatograph. The function scheduling should be
similar to the following:
PROCESS STREAM METHOD TABLE
Method Table #01
Time Function Value
0001 Temp. Contrl One
0128 Auto Zero N/A
0130 Skp, Nxt,Str= Nine
0132 Valve On One
0155 Valve Off One
0165 Baseline Proj Forward
0172 Forced Gate On @ Valley
0175 Component One
0178 Forced Gate Off @ Valley
0179 Forced Gate On @ Valley
0182 Component Two
0185 Forced Gate Off @ Valley
0186 Forced Gate On @ Valley
0188 Component Three
0191 Forced Gate Off @ Valley
0205 Forced Gate On @ Valley
0208 Component Four
0211 Forced Gate Off @ Valley
0212 Forced Gate On @ Valley
0214 Component Five
0220 Forced Gate Off @ Valley
0244 Forced Gate On @ Valley
0247 Component Six
0250 Forced Gate Off @ Valley
0259 Forced Gate On @ Valley
0262 Component Seven
0265 Forced Gate Off @ Valley
0287 Forced Gate On @ Valley
0290 Component Eight
0293 Forced Gate Off @ Valley
0710 Stream Step N/A
0768 Component Nine
0769 Component Ten
0770 Component Eleven
0771 Component Twelve
0772 Component Thirteen
0773 Component Fourteen
0774 Component Fifteen
0775 Component Sixteen
0776 Component Seventeen
0777 Component Eighteen
0778 Component Nineteen
0779 Component Twenty

Table of Contents