Page #243 background image
2000-OSM, F1 7-29
CALIBRATION STREAM METHOD TABLE
Method Table #10
Time Function Value
0001 Temp. Contrl One
0120 Noise Calc On
0125 Noise Calc Off
0128 Auto Zero N/A
0132 Valve On One
0155 Valve Off One
0160 Slope Detect On Now
0160 Peak Width Eight
0175 Component One
0175 Component One
0182 Component Two
0188 Component Three
0208 Component Four
0217 Component Five
0217 Baseline Force @ Valley
0247 Component Six
0262 Component Seven
0290 Component Eight
0290 Baseline Force @ Valley
0312 Component Nine
0312 Baseline Force @ Valley
0343 Component Ten
0343 Baseline Force @ Valley
0359 Component Eleven
0359 Baseline Force @ Valley
0390 Component Twelve
0390 Baseline Force @ Valley
0424 Component Thirteen
0424 Baseline Force @ Valley
0440 Component Fourteen
0440 Baseline Force @ Valley
0462 Component Fifteen
0462 Baseline Force @ Valley
0509 Component Sixteen
0509 Baseline Force @ Valley
0540 Component Seventeen
0540 Baseline Force @ Valley
0570 Component Eighteen
0570 Baseline Force @ Valley
0601 Component Nineteen
0601 Baseline Force @ Valley
0650 Slope Detect Off
0710 Stream Step N/A
The various time coded functions are defined as follows. Section 4 contains more complete
descriptions of these functions.
0001 Temp. Contrl One
This function ensures that the temperature of the isothermal oven is at the set point value prior to
sample injection.