2000-OSM, F1 6-19
CAUTION
See "AIR PURGING" in Section 4 for the override function. Do not perform
override until you have read Section 4 completely and you understand and can
perform the procedure requirements.
Install a temperature monitor and verify ambient temperature. Ensure air flow into the Controller
housing.
Set Point Unacceptable
Symptom
: Alarm Condition.
Any one of the following alarm messages or conditions could be present:
On Background Screen ***ALARM:*** (flashing)
On Manual Temperature Zone (#) (flashing)
Control Mode Screen *Temp Bd Alarm* (flashing)
On Alarm Screen Temperature Board:
*Zone Specific
*** TEMP ZONE (#) Alarm ***
*Set Point Failed
To modify the temperature configuration table, refer to the "DIGITAL TEMPERATURE
CONTROLLER" in Section 4.
Reconfiguration Failure
Symptom
: Alarm Condition.
Any one of the following alarm messages or conditions could be present:
On Background Screen ***ALARM:*** (flashing)
On Manual Temperature Zone (#) (flashing)
Control Mode Screen *Temp Bd Alarm* (flashing)
On Alarm Screen Temperature Board:
*Zone Specific
*** TEMP ZONE (#) Alarm ***
*Reconfiguration Failed
Remedy
: Perform the temperature board reconfiguration procedure. Be certain to hold down the
switch and respond by pressing “Y”. To modify the temperature configuration table, refer to the
"DIGITAL TEMPERATURE CONTROLLER" in Section 4.
Out of Control
CAUTION
It is the customer’s responsibility to ensure that the area is safe and and hazard-
free, and will remain so the entire time the analyzer is open. This responsibility
includes ensuring adequate ventilation in analyzer shelter and obtaining proper
work permits, etc.