8-52 2000-OSM, F1
6. On the Pres and Temp Control screen (see Figure 8-24), cursor to PRES CONTROL and press
the F2 (Manual Pressure Control ) soft key.
***** Pres and Temp Control *****
TEMP CONTROL
PRES CONTROL
TEMP CONFIGURATION
Manual Pressure View
Escape Pressure Config Alarms
Control
F1 F2 F3 F4
Figure 8-24. PRESSURE AND TEMPERATURE CONTROL SCREEN
7. On the Manual Pressure control Mode screen (see Figure 8-25), enter the zone number of the
zone you want to calibrate.
**** Manual Pressure Control Mode ****
Zone Config Actual SP Rate
COLUMN ISO 0.0241 0.0000 0.0000
ANALYSIS ISO 0.360 0.0000 0.0000
SELECTOR ISO 0.699 0.0000 0.0000
ANALYSIS ISO 0.0000 0.0000 0.0000
COLUMN PROG 0.00167 0.0000 1.0000
Valve # : [0] Time : 0125
Zone # : 1 State : Idle
Set Point : 0.00000000
Ramp Rate : 1.0000000 Ver A
VALVES : --3-5-------
Accept Calibrate
Escape Control Zone
Point
F1 F2 F3 F4
Figure 8-25. MANUAL PRESSURE CONTROL MODE SCREEN