DD+DIS150.06E
Repair and Service
Electrical and mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1 CR 30-X Chapter 3.4 / 5
04-2008 Type 5175 / 100/110 Agfa Company Confidential
LED Meaning
/Reset Orange LED, default OFF. ON during board reset.
/M32_Reserve2 Green LED, default OFF. For R&D purpose only.*
/M32_Reserve1 Green LED, default ON. For R&D purpose only.*
/FPGA_Test3 Green LED, default ON. For R&D purpose only.*
/FPGA_Test4 Green LED, default OFF. For R&D purpose only.*
/CS3 Green LED, default ON. For R&D purpose only.*
/CS2 Green LED, default OFF. For R&D purpose only.*
/CS1 Green LED, default OFF. For R&D purpose only.*
/CS0 Green LED, default OFF. For R&D purpose only.*
+5 V PMT Green LED, default ON. Voltage +5 V okay. Only
available on PMI2 board.
+8 V DISP Green LED, default ON. Voltage +8 V for display okay.
+12 V CAN Green LED, default ON. Voltage +12 V for CAN bus
okay.
+8 V Analog Green LED, default ON. Voltage +8 V for analog circuit
okay.
-10 V PMT Green LED, default ON. Voltage -10 V for PMT okay.
+20 V PMT Green LED, default ON. Voltage +20 V for PMT okay.
-12 V CAL Green LED, default ON. Voltage - 12 V for calibration
board okay.
+12 V CAL Green LED, default ON. Voltage +12 V for calibration
board okay.
-12 V OPT Green LED, default ON. Voltage -12 V for optic module
okay.
+12 V OPT Green LED, default ON. Voltage +12 V for optic
module okay.
+24 V POL Green LED, default ON. Voltage +24 V for polygon
okay.
+3,3 V Green LED, default ON. Voltage +3,3 V okay.
FWI_Orange Orange LED, default ON. FWI cable connected
between digitizer and processing station.
FWI_Green Green LED, default ON. Voltage for FireWire Interface
okay.
*If these LEDs are off, this does not necessarily mean malfunction of the board. May depend on SW version,
how LED is programmed.