6.1.3 Creating the Flatfield Test Image
(1) Mount the Cu Filter at the General Radiology modality.
(2) Expose the prepared cassette with approximately 10 µGy.
(3) Turn the cassette by 180.
(4) Expose again with 10 µGy.
NOTE:
Select the following exposure parameters using a 1.5 mm Cu filter
(CM+9 5155 1015 2) to obtain a dose of 10 µGy:
12 mAs
75 kVp
1.3 m (51,2”) distance and
Large focus
The values are approximate and may vary among the X-ray devices.
IMPORTANT:
For exposure place the cassette in length direction to the X-ray tube.
OK Not OK
Figure 20
The entire image plate must be fully exposed!
The collimated field must be larger than the image plate!
Image Plate
E
xpose
d
area
Image Plate
OK Not OK
Figure 21
Edition 1, Revision 5 CR 30-X Chapter 1 / 19
11-2009 Type 5175 / 100/110 Agfa Company Confidential