DD+DIS150
.06E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4 CR 30-X Chapter 3.8 / 24
01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
13 Software Release CRE_1901
Type of new Software
Software Upgrade
Release Date
12/2010
Digitizer SNs
Type 5175/100 SN ≥ 31.684, additionally in SN 31.680
Type 5175/105 SN ≥ 39.001
Type 5175/110 SN ≥ 40.342
Type 5175/120 SN ≥ 41.001
Type 5175/125 SN ≥ 43.001
Availability
MEDNET, GSO Library* (25 MB size)
Prerequisites
Software CRE_1901 has only been released with
processing station software NX 2.0.6805 SU1 and SU2.
Together with software upgrade CRE_1901 it is
mandatory to upgrade to NX 2.0.6805 SU1 or SU2, too, if
not yet done earlier.
Reference Document
DD+DIS308.10E, Service Bulletin No. 44, Software
CRE_1901 and new Digitizer Subtypes available
(Intranet Link
/ Extranet Link)
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
13.1 New Features of Software CRE_1901
Support of new digitizers sub types 105, 120 and 125 with different throughput
values for different markets. For details refer to service documentation chapter 11,
Installation Planning.
The Service Client can be installed on Windows 7, 32 bit systems additionally to
Windows XP SP2
Improved service functionalities:
o The serial number of modules is not displayed anymore in the service client
menu “Exchanged components”.
Reason: Some modules have the correct serial number “0”, what was
interpreted as a wrong serial number display.
o Improved stall calibration