Functional Description
DD+DIS150.06E
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
4 Digitizer Boot-Up
Digitizer
Switch-On
Digitizer idle
Audible
Visible
Send Self Test
Signal to all SW modules
Error message
in NX Error Viewer.
Boot-up stops
Error message
in NX Error Viewer.
Boot-up stops
Wait till the NX
sends "enable feeder"
message (LED blinks red)
Boot-up of PMI
and all other CAN nodes
Warning message
in Error Viewer
Remove Cassette
or IP and continue
IP Clean-up
Cycle
Initialize all Modules
= set all motors to HOME
repeat endless
if required
NX connected?
Selftest okay?
Initialization okay?
NX Viewer
running?
Cassette or IP
found?
NO
NO
NO
NO
YES
YES
YES
NO
YES
YES
No audible
activities
Approx. 10 s.
No audible
activities Motors audible
CR30-X Boot-Up
LED
green on
LED dark
LED red blinking
Polygon audible
3 x
beep
3
2
1
3
3 x beep indicates: PMI + CAN nodes are started up.
Legend:
1
This selftest is without motor actvities (except polygon).
It checks e.g. whether motors and sensors are connected.
SW Modules:
517502eb.cdr
- Image Capture Control
- Cassette Handling Control
- IP Handling Control
- Slow Scan Control
Connection only possible if
- ASAP service on NX started.
- FireWire cable connected
2
Approx. 10 s.
Approx. 50 s.
Approx. 1 s.
3
Figure 6
Chapter 2 / 6 CR30-X Edition 1, Revision 0
Agfa Company Confidential Type 5175/100