DD+DIS238.06E
Generic Safety Directions
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 26 of 28
Secured screws:
CAUTION:
Opening screws secured by red lacquer may misalign important device
adjustments:
Do not open screws that are secured by red lacquer.
Opening PCs and Workstations:
Warning:
Electrical shock and damage to the equipment possible.
Only open the PC or workstation if explicitly stated in the service manual.
Unplug before opening.
Observe anti-static safety regulations.
Replacing fuses:
Warning:
Replacing fuses by wrong type may lead to fire hazard!
Use only fuses of the exact value and characteristics stated in the service manual or on
the device.
20 Safety Directions for remote Service Activities
Remote Service Activities:
Warning:
During remote service activities images can be lost.
Inform the customer prior to remote service activities to finish the current work and to
stop working on the system.