DD+DIS150
.06E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3 Software Patch CRE_1206C
Type of new Software
Software Upgrade
Release Date
01/2007
Digitizer SNs
1021
Availability
MEDNET, GSO Library* (600 kB size)
Prerequisites
Software CRE_1206 has to be installed before installing
the software patch CRE_1206C.
Reference Document
DD+DIS001.07E, Service Bulletin No. 05, Displayed
LgM-Values too high: Install Software Patch CRE_1206C
(Intranet Link / Extranet Link)
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
3.1 Solved Bugs with Software CRE_1206C compared to CRE_1206A
Solved wrong image plate sensitivity which was sent from the digitizer to the NX
workstation resulting in too high displayed lgM-values.
Edition 1, Revision 4 CR 30-X Chapter 3.8 / 7
01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential