DD+DIS150
.06E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
The software version of the digitizer and that
one installed on PC do not match.
Please contact service to finish installation.
(Error 2712)
x
517501ax.cdr
NOTE:
If error 2712 pops up on the control PC:
This error indicates, that the installed
software in the Digitizer and on the Control
PC (ASAP software) are not identical.
This error will be solved with steps
described in section 5.2.
Figure 10
5 Configuring Digitizer Settings
The purpose of this section is to show the required installation tasks.
The following table lists these tasks:
Task Required
Time
Importance Task Code* Section
Fill in site specific data. 5 minutes mandatory INS 5.1
Check software version. 2 minutes mandatory INS 5.2
Configure Maintenance
Indicator.
5 minutes optional INS 5.3
*Complete required time per Task Code to be entered in the Agfa repair report.
Edition 1, Revision 5 CR 30-X Chapter 1 / 12
11-2009 Type 5175 / 100/110 Agfa Company Confidential