DD+DIS238.06E
Generic Safety Directions
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3 Generic Safety Directions for HealthCare Imaging Products Page 20 of 28
17 Safety Directions for Installation Planning Activities
Protecting CR (Computed Radiography) Equipment against scattered X-Rays:
Warning:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this way,
that the annual dose equivalent at the installation place will not exceed 1 mSv.
Protecting Film-Screen Systems against scattered X-Rays:
Warning:
Film is sensitive for X-rays. Poor image quality possible.
The film-screen system shall be protected against X-ray radiation this way, that the annual
dose equivalent at the installation place will not exceed 1 mSv.
Accessibility of the power disconnection device:
Warning:
Electrical device. Shock possible.
INSTRUCTION:
Do not position Agfa EQUIPMENT so it is difficult to operate the disconnection device
when an APPLIANCE COUPLER or separable plug is used as isolation.
Local and International wiring regulations must be observed. Check all supplies and
voltages, currents, trips and fuses with the Hospital facilities department or their
engineers.