DD+DIS150
.06E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
2 Adjustments
2.1 Drawer Unit Adjustment
Purpose
Adjustment of the optimum height and lateral position of the upper drawer unit plate
for reliable image plate transport in and out of the cassette.
Use Case
After replacement of the drawer unit
REQUIRED TOOLS:
3 mm
Drawer adjustment tool
NOTE:
Two versions of the drawer adjustment tool are available:
m
i
n
.
ma
x
.
Version 1, delivered
till October 2009
Version 2, delivered
as of October 2009
Figure 1
Difference: Drawer adjustment tool version 2 has an additional gauge integrated to
check the maximum height position (marked with text "max.").
The drawer adjustment tool is part of delivery of following spare parts:
Drawer Unit
Drawer Motor
Its is also available from spare parts stock with spare part number*: CM+9 5175 6001 0
* The last digit in the spare part number indicates the spare part revision at release of
this document. When ordering, the actual revision of the spare part is delivered.
This document describes the adjustment with drawer adjustment tool version 2.
Where applicable, differences to version 1 are described.
REQUIRED TIME:
30 minutes
Edition 1, Revision 3 CR 30-X Chapter 3.6 / 5
10-2009 Type 5175 / 100/110 Agfa Company Confidential