Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
4 Software Patch CRE_1206D
Type of new Software
Software Update
Release Date
02/2007
Digitizer SNs
1366 plus 1187, 1260, 1261, 1263, 1338, 1355, 1357,
1363
Availability
MEDNET, GSO Library* (500 kB size)
Prerequisites
Software CRE_1206 has to be installed before installing
the software patch CRE_1206D.
Reference Document
DD+DIS027.07E, Service Bulletin No. 06, Interfering
Noise in the Image:
Install Software Patch CRE_1206D (obsolete Document)
*Note:
On MedNet, GSO Library always the latest available software release can be
downloaded.
4.1 Solved Bugs with Software CRE_1206D compared to CRE_1206C
Solved insufficient initialization of the RAM on the PMI board, what caused interfering
noise in the image.
Edition 1, Revision 4 CR 30-X Chapter 3.8 / 8
01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential