Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4 CR 30-X Chapter 3.8 / 16
01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
8.2.4 Additional, minor Bug Fixes
Following a list of minor software bugs that appeared at internal tests and are solved
with CRE_1509:
Solved Service Function related Software Bugs:
In some cases, when working with the Service Client, the error viewer was not
visible, if an error occurred (e.g. in diagnose cycle). To solve the problem in the
worst case a reset of the control PC was required. In SW 1509 always the error
viewer is visible if an error occurs while working with the service client.
Total number of cycles of the product was not written to the E-label and the
infocounter.
Warnings were not stored in infocounter “error history”.
Stall Calibration details were not logged in Service Activity Log file
After software installation the message “SW upload required appeared.
It was not possible to perform the stall calibration remotely.
In some cases the service client could not re-establish communication after
software upload. It was required to reboot the control PC.
Retries during IP handling are not logged in the infocounter.
Software upload gave (false) error 15399.
Error messages were not displayed when performing a diagnose cycle.
Solved Error Handling Software Bugs:
In case of a Photomultiplier defect the digitizer software crashed: No error was
displayed.
If the digitizer had problems writing to the cassette chip, not only NX message
“119: The file could not be saved” appeared, but also a false error message
“Problem during scan: The image may not have the required quality”.
If FireWire communication was interrupted during scan, communication could only
be re-established by a reset of both, digitizer and control PC.
During boot-up the stepper motors did not retry to go to home position if a problem
occurred. Now three retries are performed.
An aging photomultiplier could not be detected.
Retry sequence of cassette opener did not work, if cassette cover was
mechanically blocked.
Some timeouts during scan were not transferred to the control PC, e.g. if “scan
command” did not return a response.