DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4 CR 30-X Chapter 3.8 / 5
01-2011 Type 5175/100/105/110/120/125 Agfa Company Confidential
1 Software Release CRE_1010A
Type of new Software
1st customer release
Release Date
09/2006
Digitizer SNs
1001
Availability
MEDNET, GSO Library* (8 MB size)
Prerequisites
None
Reference Document
DD+DIS365.06E, Service Bulletin No. 01, CR 30-X
System released for Sales (Intranet Link / Extranet Link)
1.1 Supported Functions
All functions as specified:
Open the inserted cassette (15 x 30 cm; 18 x 24 cm; 24 x 30 cm; 35 x 43 cm) and
transport the IP to the Scan Unit.
Acquire the ID data via fast ID mode from the processing station.
Scan (= stimulate and scan) the image on the IP.
Send the raw image to the processing station.
Erase the IP.
Put the IP back in its cassette, close cassette.
Update ID data on IP (cycle counter).
1.2 Limitations
Shading calibration curves are stored at the processing station only. This means, if a
digitizer is operated with another processing station without restore, its shading
calibration is lost.