DD+DIS150.06E
Repair and Service
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3 Safety Notes for Service Activities
3.1 Safety Notes for Installation Planning
Warning:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this
way, that the annual dose equivalent at the installation place will not exceed 1 mSv.
WARNING:
Excessive vibrations during scanning may decrease image quality.
The structure and stability of the table used, need to be suitable in relation with the size
and weight of the system. The table may not be subject to excessive shock and
vibrations from other sources.
WARNING:
The digitizer uses high voltage. Electrical shock possible.
A ground fault interrupter (GFI) (I
N
= 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
Ensure that there is either a mains plug or an all-cable disconnecting device in the
internal installation fitted near the digitizer and that it is easily accessible.
WARNING:
Vibrations caused by transport in a truck (mobile usage) or earthquakes may
cause that the digitizer falls off the table.
During transport in a truck for mobile usage and at installations in earthquake areas the
digitizer has to be secured by the optional Wall Fixation Kit. See chapter 11, Installation
Planning, section "Accessories".
Edition 1, Revision 1 CR 30-X Chapter 3.1 / 8