DD+DIS150.06E
Repair and Service
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 CR 30-X Chapter 3.5 / 12
03-2009 Type 5175 / 100/110 Agfa Company Confidential
(5) Remove 3 screws of the top cover.
(6) Remove top cover.
Figure 18
ASSEMBLY:
(1) Mount in reverse order.
Verification:
(1) Switch on the digitizer.
(2) Wait till the green light at the user interface lights up
(takes approx. 1,5 min.).
Result:
The top cover is replaced.