Performance Test
E5071C Performance Test Record
2. Performance Test
Direction: S34 (Magnitude)
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[dB rms]
Test result
[dB rms]
Measurement
uncertainty [dB]
9 k - 30 k 3 k < 0.004 0.00009
30 k - 10 M 3 k < 0.003 0.00009
10 M - 3 G 70 k < 0.004 0.00033
3 G - 4.38 G 70 k < 0.004 0.00033
4.38 G - 8.5 G 70 k < 0.006 0.00023
Direction: S34 (Phase)
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[deg rms]
Test result
[deg rms]
Measurement
Uncertaintiy [deg]
9 k - 30 k 3 k < 0.035 0.00077
30 k - 10 M 3 k < 0.02 0.0006
10 M - 3 G 70 k < 0.035 0.001
3 G - 4.38 G 70 k < 0.035 0.001
4.38 G - 8.5 G 70 k < 0.05 0.0023
Option 485
Direction: S11 (Magnitude)
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[dB rms]
Test result
[dB rms]
Measurement
uncertainty [dB]
100 k - 10 M 3 k < 0.003 0.00009
10 M - 3 G 70 k < 0.004 0.00033
3 G - 4.38 G 70 k < 0.004 0.00033
4.38 G - 8.5 G 70 k < 0.006 0.00023
Direction: S11 (Phase)
4.38 G - 8.5 G 70 k < 0.05 0.0023
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[deg rms]
Test result
[deg rms]
Measurement
Uncertaintiy [deg]
Frequency
Range[Hz]
IF bandwidth
[Hz]
Test limit
[deg rms]
Test result
[deg rms]
Measurement
Uncertaintiy [deg]
100 k - 10 M 3 k < 0.02 0.0006

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals