288 Chapter 6
Replacement Procedure
USB (USBTMC) Interface Card Replacement
USB (USBTMC) Interface Card Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-14 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266.
Step 2. Remove the three TORX T10 screw (item 1) fastening the angle and lift it.
Step 3. Disconnect the flat cable (item 2) from the PCI DSP card.
Step 4. Remove the two TORX T10 screws (item 3) fastening the card to the rear panel.
Step 5. Disconnect the assembly (item 4) from the Digital Motherboard and lift it.
Step 6. Remove the two TORX T10 screws (item 5) fastening the angle.