308 Chapter 6
Replacement Procedure
Fan Replacement
Fan Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
A fine stick such as Hex key, 1.5 mm
Cutting plier or scissors
Removal Procedure
Refer to Figure 6-24 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266.
Step 2. Remove the removal hard disk assembly as described in “Removable Hard Disk Assembly
Removal” on page 268.
Step 3. Disconnect the fan cables (item 1) from the analog base module.
Step 4. Disconnect the flat cable (item 2) from the analog base module.
Step 5. Disconnect the flat cable (item 3) from the digital bridge board.
Step 6. Remove the six TORX T10 screws (item 4) fastening the fan plate.
Step 7. Cut the cable tie (item 5).
Step 8. Remove the twelve rivet (plastic)(item 6) fastening the fan.
Step 9. Remove the fan (item 7) from fan plate.