Chapter 2 33
Performance Test
4. RF OUTPUT LEVEL LINEARITY TEST
2. Performance Test
4. RF OUTPUT LEVEL LINEARITY TEST
Description
This test checks the level accuracy of the E5071C test port output signal across the
specified level range. The RF output level is measured for power level settings of -20 dBm
to 10 dBm in 2.5dB step increments at 9 kHz, 100 kHz, 50 MHz, 3 GHz, 4.5 GHz, 5 GHz,
and -20 dBm to 9 dBm in 2.5 dB step increments at 6 GHz, and -20 dBm to 8 dBm in 2.5
dB step increments at 7 GHz, and -20 dBm to 7 dBm in 2.5 dB step increments at 8.5 GHz.
Test equipment
Power meter Agilent E4419A/B
Power sensor Agilent E9304A Opt. H18 or Agilent E9304A Opt.
H19